Thursday 16 May 2013

ATOMIC FORCE MICROSCOPY


Atomic force microscopy (AFM) or scanning force microscopy (SFM) is a very high-resolution type of scanning probe microscopy, with demonstrated resolution on the order of fractions of a nanometer, AFM provides a 3D profile of the surface on a nanoscale, by measuring forces between a sharp probe (<10 nm) and surface at very short distance (0.2-10 nm probe-sample separation). The probe is supported on a flexible cantilever. The AFM tip “gently” touches the surface and records the small force between the probe and the surface.
The AFM measures the forces acting between a fine tip and a sample. The tip is attached to the free end of a cantilever, The cantilever is typically silicon or silicon nitride with a tip radius of curvature on the order of nanometres and it is brought very close to a surface. Attractive or repulsive forces resulting from interactions between the tip and the surface will cause a positive or negative bending of the cantilever. The bending is detected by means of a laser beam, which is reflected from the back side of the cantilever. The figure  shows the basic concept of AFM :


What types of forces are measured?

Plot of force as a function of probe-sample separation
The dominant interactions at short probe-sample distances in the AFM are Van der Waals interactions. However long-range interactions (i.e. capillary, electrostatic, magnetic) are significant further away from the surface. These are important in other SPM methods of analysis. During contact with the sample, the probe predominately experiences repulsive Van der Waals forces (contact mode). This leads to the tip deflection described previously. As the tip moves further away from the surface attractive Van der Waals forces are dominant (non-contact mode).


Modes of operation

CONTACT MODE

In the So-called contact-AFM mode, the tip makes soft “physical contact” with the surface of the sample. The deflection of the cantilever Dx is proportional to the force acting on the tip, via Hook’s law, F=-k. x, where k is the spring constant of the cantilever. In contact-mode the tip either scans at a constant small height above the surface or under the conditions of a constant force. In the constant height mode the height of the tip is fixed, whereas in the constant-force mode the deflection of the cantilever is fixed and the motion of the scanner in z-direction is recorded. By using contact-mode AFM, even “atomic resolution” images are obtained. For contact mode AFM imaging, it is necessary to have a cantilever which is soft enough to be deflected by very small forces and has a high enough resonant frequency to not be susceptible to vibration instabilities. Silicon Nitride tips are
used for contact mode. In these tips, there are 4 cantilever with different geometries attached to each substrate, resulting in 4 different spring constants.
Probe with four different cantilevers with different spring constants (N/m
Advantages:
  • High scan speeds
  • Atomic resolution” is possible
  • Easier scanning of rough samples with extreme changes in vertical topography
Disadvantages:
  • Lateral forces can distort the image.
  • Capillary forces from a fluid layer can cause large forces normal to the tip sample interaction
  • Combination of these forces reduces spatial resolution and can cause damage to soft samples.
INTERMITTENT MODE (TAPPING)


The imaging is similar to contact. However, in this mode the cantilever is oscillated at its resonant frequency, Figure 4. The probe lightly “taps” on the sample surface during scanning, contacting the surface at the bottom of its swing. By maintaining a constant oscillation amplitude a constant tip-sample interaction is  maintained and an image of the surface is obtained.

Advantages:

  • Higher lateral resolution (1 nm to 5 nm).
  • Lower forces and less damage to soft samples in air.
  • Almost no lateral forces.
Disadvantage:

  • Slower scan speed than in contact mode.
  
  NON CONTACT MODE

(Attractive VdW) The probe does not contact the sample surface, but oscillates above the adsorbed fluid layer on the surface during scanning. (Note: all samples unless in a controlled UHV or environmental chamber have some liquid adsorbed on the surface). Using a feedback loop to monitor changes in the amplitude due to attractive VdW forces the surface topography can be measured.

Advantage:

  • VERY low force exerted on the sample(10-12 N), extended probe lifetime.
Disadvantages:

  • Lower lateral resolution, limited by tip-sample separation.
  • Contaminant layer on surface can interfere with oscillation; usually need ultra-high vacuum (UHV) to have best imaging.
  • Slower scan speed to avoid contact with fluid layer.
  • Usually only applicable in extremely hydrophobic samples with a minimal fluid layer.


Applications :

The number of applications for AFM has exploded since it was invented in 1986 and Nowadays this technique is involved in many fields of Nanoscience and nanotechnology. The remarkable feature of STM and AFM instruments is their ability to examine samples not only in an ultrahigh vacuum but also on ambient conditions or even in liquids. AFM can image the non-conducting surfaces, and therefore it is very suitable for biological systems.

Possible applications of AFM are :

  • Substrate roughness analysis.
  • Step formation in thin film epitaxial deposition.
  • Pinholes formation or other defects in oxides growth.
  • Grain size analysis.
  • Phase mode is very sensitive to variations in material properties, including surface stiffness, elasticity and adhesion.
  • Comparing the tip-samples forces curves for materials to study the ratio of Young´s Modulus (graphite as a reference for measure of the indentation).
  • Obtaining information of what is happening under indentation at very small loads .
  • By In situ AFM analysis with changes in temperature we can study changes in the structure.


References :

1.  Basic Theory Atomic Force Microscopy (AFM) by  Robert A. Wilson and Heather A. Bullen,* Department of Chemistry, Northern Kentucky University, Highland Heights
2.  Principles of atomic force microscopy (AFM) written by Arantxa Vilalta-Clemente , Aristotle University, Thessaloniki, Greece and Kathrin Gloystein, Aristotle University, Thessaloniki, Greece





4 comments:

  1. Amazing blog. Thanks for sharing this wonderful information with us
    Atomic Force Microscope

    ReplyDelete
  2. Thanks so much for a great post. I'd like to know more about these topics and hope that I can receive more insight into this topic.
    For more information visit here: Atomic Force Microscope

    ReplyDelete
  3. Hey, Nice blog and knowledgeable content.
    Atomic Force Microscope

    ReplyDelete