WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE (WD-XRF)

INTRODUCTION

Wavelength Dispersive X-Ray Fluorescence Spectrometry (WD-XRF) is the oldest method of measurement of X-rays, introduced commercially in the 1950’s. This name is descriptive in that the radiation emitted from the sample is collimated with a Soller collimator, and then impinges upon an analyzing crystal. The crystal diffracts the radiation to different extents, according to Bragg’s law, depending upon the wavelength or energy of the X radiation. This angular dispersion of the radiation permits the sequential or simultaneous detection of X-rays emitted by elements in the sample. Simultaneous instruments normally contain several sets of analyzing crystals and detectors; one is adjusted for each desired analyte in the sample. These instruments tend to be very expensive, but efficient for the routine determination or preselected elements.

WD-XRF spectrometers are usually larger and more expensive than other spectrometers. Because the analyzing crystal d-spacing determines wavelength sensitivity, they are usually more sensitive than other spectrometers. To overcome losses in X-ray optics of the WD-XRF spectrometers and to maximize primary radiation intensity, X-ray tubes are usually employed. The sample is usually held under vacuum to reduce contamination and avoid absorption of light element characteristic radiation in air.

Typical uses of WD-XRF include the analysis of oils and fuel, plastics, rubber and textiles, pharmaceutical products, foodstuffs, cosmetics and body care products, fertilizers, minerals, ores, rocks, sands, slags, cements, heat-resistant materials glass, ceramics, semiconductor wafers; the determination of coatings on paper, film, polyester and metals; the sorting or compositional analysis of metal alloys, glass and polymeric materials; and the monitoring of soil contamination, solid waste, effluent, cleaning fluids, sediments and air filters.

PRINCIPLE OF  WD-XRF

WD-XRF spectrometers measure X-ray intensity as a function of wavelength. This is done by passing radiation emanating from the specimen through an analyzing diffraction crystal mounted on a 2θ goniometer. By Bragg’s Law, the angle between the sample and detector yields the wavelength of the radiation:

2sin θ = nλ ;

where:
d    is the d-spacing of the analyzing crystal,
θ    is half the angle between the detector and the sample,
n    is the order of diffraction.

The analyzing crystal must be oriented so that the crystal diffraction plane is directed in the appropriate direction. Figure shows a simplified schematic of the WD-XRF spectrometer.  A scintillation or flow-proportional detector usually measures the fluoresced radiation. The heights of the resulting pulses are proportional to energy so using a pulse height analyzer (PHA), scattered or undesired diffraction-order X-rays can be ejected. The X-ray beam is usually collimated before and after the analyzing crystal. Each of the components showed in the below (Figure 1)  were be described in the following sections.

1. Schematic description of WD-XRF principle


1. COLLIMATOR MASKS

The collimator masks are situated between the sample and collimator and serve the
purpose of cutting out the radiation coming from the edge of the cup aperture (Figure 2).

The size of the mask is generally adapted to suit of the cup aperture being used.
The masks perform one of the two functions: background reduction and improved

fluorescence (Figure 3).

2.Use of Cu 200 μm filter for cutting off the radiation coming from Rh X-ray tube.

3. Use of Al 100 μm filter for improvement of the ratio peak/background.

2. COLLIMATOR

4. Collimators with different angles of resolution.
Collimators consist of a row of parallel slats (Figure 4) and select a parallel beam of X-rays coming from the sample and striking the crystal. The spaces between the slats determine the degree of parallelism and thus the angle resolution of the collimator.
A 0.077° collimator is adequate for high resolution measurement parameters. Collimators with low resolution (e.g. 1.5 -2.0°) are advantageous for light elements such as Be, B and C (Figure 5). Using a collimator with a low resolution increases then intensity significantly. This enables intensity to be increased without a loss in angle resolution when analyzing light elements.

5. Example of the influence of collimator resolution on the intensity of a light
element.


3. THE ANALYZING CRYSTALS

a. Bragg’s Law

Crystals consist of a periodic arrangement of atoms (molecules) that form the crystal lattice. In such an arrangement of particles you generally find numerous planes running in different directions through the lattice points (= atoms, molecules), and not only horizontally and vertically but also diagonally. These are called lattice planes. All of the planes parallel to a lattice plane are also lattice planes and are at a defined distance from each other. This distance is called the lattice plane distance “d”.

6. Bragg’s Law.
When parallel X-ray light strikes a lattice plane, every particle within it acts as a scattering center and emits a secondary wave. All of the secondary waves combine to form a reflected wave. The same occurs on the parallel lattice planes for only very little of the X-ray wave is absorbed within the lattice plane distance “d”. All these reflected waves interfere with each other. If the amplification condition “phase difference = a whole multiple of the wavelength” (Δλ = nλ) is not precisely met, the reflected wave will interfere such that cancellation occurs. All that remains is the wavelength for which the amplification condition is met precisely. For a defined wavelength and a defined lattice plane distance, this is only given with a specific angle, the Bragg angle (Figure 6).



Under amplification conditions, parallel, coherent X-ray light (1,2) falls on a crystal
with a lattice plane distanced ‘d’ and is scattered below the angle θ (1,2). The proportion of the beam that is scattered on the second plane has a difference of ‘ACB’ to the proportion of the beam that was scattered at the first plane. The amplification condition is fulfilled when the phase difference is a whole multiple of the wavelength λ. This results in Bragg’s Law:

2sin θ = nλ ;

n = 1, 2, 3… Reflection order.

On the basis of Bragg’s Law, by measuring the angle θ, we can determine either the
wavelength λ, and thus chemical elements, if the lattice plane distance ‘d’ is known or, if the wavelength λ is known, the lattice plane –value distance ‘d’ and thus the crystalline structure.
This provides the basis for two measuring techniques for the quantitative and qualitative determination of chemical elements (XRF) and crystalline structures (molecules, XRD), depending on whether the wavelength λ or the 2d-value is identified by measuring the angle θ.
  • In X-ray diffraction (XRD) the sample is excited with monochromatic radiation of a known wavelength (λ) in order to evaluate the lattice plane distance as per Bragg’s equation.
  • In XRF, the ‘d’-value of the analyzer crystal is known and we can solve Bragg’s equation for the element characteristic wavelength (λ).

b.  Reflections of Higher Orders


Figures 7a and 7b illustrate the reflections of the first and second order of one wavelength below the different angles θ1 and θ2. Here, the total reflection is made up of the various reflection orders (1, 2 …, n). The higher the reflection order, the lower the intensity of the reflected proportion of radiation generally is. How great the maximum detectable order is depends on the wavelength, the type of crystal used and the angular range of the spectrometer.

7 a.  First order reflection: λ = 2 d sin θ1.

7 b.  Second order reflection: 2λ = 2 d sin θ2.


It can be seen from Bragg’s equation that the product of reflection orders ‘n = 1; 2; ..’
and wavelength ‘λ’ for greater orders, and shorter wavelengths ‘λ* < λ’ that satisfy the
condition ‘λ* = λ/n’, give the same result.

Accordingly, radiation with one half, one third, one quarter etc. of the appropriate
wavelength (using the same type of the crystal) is reflected below the identical angle θ:

1λ = 2(λ / 2) 3(λ / 3) 4(λ / 4) = ……......

As the radiation with one half of the wavelength has twice the energy, the radiation with one third of the wavelength three times the energy etc., peaks of twice, three times the energy etc. can occur in the pulse height spectrum (= energy spectrum) as long as appropriate radiation sources.

c.  Crystal Types

The wavelength dispersive X-ray fluorescence technique can detect every element
above the atomic number 4 (Be). The wavelengths cover the range of values of four
magnitudes: 0.01 – 11.3 nm. As the angle θ can theoretically only be between 0° and 90° (in practice 2° to 75°), sinθ an only accept values between 0 and +1. When Bragg’s equation is applied:

  0 < nλ/2d = Sinθ < 1

 This means that the detectable element range is limited for a crystal with a lattice plane difference ‘d’. Therefore a variety of crystal type with different ‘2d’ values is necessary to detect the whole element range (from atomic number 4).

Besides the ‘2d’ values, the following selection criteria must be considered when a

particular type of crystal is to be used for a specific application:
  • Resolution
  • Reflectivity ( intensity)
Further criteria can be:
  • Temperature stability
  • Suppression of higher orders
  •       Crystal fluorescence.

d.  Dispersion, Line Separation

The extent of the change in angle Δθ upon changing the wavelength by the amount Δλ (thus: Δθ/Δλ) is called “dispersion”. The greater the dispersion, the better is the separation of two adjacent or overlapping peaks. Resolution is determined by the dispersion as well as by surface quality and the purity of the crystal.
Mathematically, the dispersion can be obtained from the differentiation of the Bragg

equation. 

Δθ /Δλ =  n/2dsinθ

It can be seen from this equation that the dispersion (or peak separation) increases as the lattice plane distance ‘d’ declines.

e. Synthetic Multilayers

8.  Diffraction in the layers (here: Si/W) of a multilayer.
Multilayers are not natural crystals but artificially produced ‘layer analyzers’. The lattice plane distances ‘d’ are produced by applying thin layers of two materials in alternation on to a substrate (Figure 8). Multilayers are characterized by high reflectivity and a somewhat reduced resolution. For the analysis of light elements the multilayer technique presents an almost revolutionary improvement for numerous applications in comparison to natural crystals with large lattice plane distances.



4.    DETECTORS

When measuring X-ray, use is made of their ability to ionize atoms and molecules, i.e. to displace electrons from their bonds by energy transference. In suitable detector materials, pulses whose strengths are proportional to the energy of the respective X-ray quants are produced by the effect of X-ray. The information about the X-ray quarts energy is contained in the registration of the pulse height. The number of X-ray quants per unit of time, e.g. pulses per second (cps = counts per second, KCps = kilocounts per second), is called their intensity and contains in a first approximation the information about the concentration of the emitting in the sample. Two main types of detectors are used in wavelength dispersive X-ray fluorescence spectrometers: the gas proportional counter and the scintillation counter.

a. Gas Proportional Counter

9: A gas proportional counter.
The gas proportional counter comprises a cylindrical metallic tube in the middle of which a thin wire (counting wire) is mounted. This tube is filled with a suitable gas (e.g. Ar+ 10% CH4). A positive high voltage (+U) is applied the wire. The tube has a lateral aperture or window that is sealed with a material permeable to X-ray quants (Figure 9).


An X-ray quant penetrates the window into the counter’s gas chamber where it is absorbed by ionizing the gas atoms and molecules. The resultant positive ions move to the cathode (tube), the free electrons to the anode, the wire. The number of electron-ion pairs created is proportional to the energy of the X-ray quant. To produce an electron-ion pair,approx. 0.03 keV are necessary, i.e. the radiation of the element boron (0.185 keV) produces approx. 6 pairs and the K-alpha radiation of molybdenum (17.5 keV) produces approx. 583 pairs. Due to the cylinder geometric arrangement, the primary electrons created in this way see an increasing electrical field on route to the wire.

The high voltage in the counting tube is now set so high that the electrons can obtain enough energy from the electrical field in the vicinity of the wire to ionize additional gas particles. An individual electron can thus create up to 10.000 secondary electron-ion pairs. The secondary ions moving towards the cathode produce measurable signal. Without this process of gas amplification, signals from boron, for example, with 6 or molybdenum with 583 pairs of charges would not be able to be measured as they would not be sufficiently discernible from the electronic noise. As amplification is adjustable via high voltage in the counting tube and is set higher for measuring boron than for measuring molybdenum. The subsequent pulse electronics supply pulses of voltage whose height depends, amongst other factors, on the energy of the X-ray quants.

b.  Scintillation Counters

10.   Scintillation counter including photomultiplier.
The scintillation counter, “SC”, used in XRF comprises a sodium iodide crystal in which thallium atoms are homogeneously distributed ‘NaI(Tl)’. The density of the crystal is sufficiently high to absorb all the XRF high energy quants. The energy of the pervading X-ray quants is transferred step by step to the crystal atoms that then radiate light and cumulatively produce a flash. The amount of light in this cintillation flash is proportional to the energy that the X-ray quant has passed to the crystal. The resulting light strikes a photocathode from which electrons can be detached very easily. These electrons are accelerated in a photomultiplier and, within an arrangement of dynodes, produce so-called secondary electrons giving a measurable signal once they have become a veritable avalanche (Figure 10). The height of the pulse of voltage produced is, as in the case of the gas proportional counter, proportional to the energy of the detected X-ray quant.



c.  Pulse Height Analysis (PHA), Pulse Height Distribution

If the number of the measured pulses (intensity) dependent on the pulse height is displayed in a graph, we have the ‘pulse height spectrum’. Synonymous terms are: ‘pulse height analysis’ or ‘pulse height distribution’. As the height of the pulses of voltage is proportional to the X-ray quants energy, it is also referred to as the energy spectrum of the counter (Figure 11a and 11b). The pulse height is given in volts, scale divisions or in ‘%’ (and could be started in keV after appropriate calibration). The “%”-scale is defined in such a way that the peak to be to be analyzed appears at 100 %.

11.a.    Pulse height distribution (S) Gas proportional counter

11.b .   Pulse height distribution (Fe) Scintillation counter.


12.    Pulse height distribution (Fe) with escape peak.
If argon is used as the counting gas component in gas proportional counters, an
additional peak, the escape peak (Figure 12), appears when X-ray energies are irradiated that are higher than the absorption edge of argon.

The escape peak arises as follows:
The incident X-ray quant passes its energy to the counting gas thereby displaying a K electron from an argon atom. The Ar atom can now emit an Ar Kα1,2 X-ray quant with an energy of 3 keV. If this Ar-fluorescence escapes from the counter then only the incident energy minus 3 keV remains for the measured signal. A second peak, the escape peak that is always 3 keV below the incident energy, appears in the pulse height distribution.

When using other counting gases (Ne, Kr, Xe) instead of argon, the escape peaks appear with an energy difference below the incident energy that is equivalent to the appropriate emitted fluorescence radiation (Kr, Xe). Using neon as the counting gas component produces no recognizable escape peak as the Ne K-radiation, with energy of 0.85 keV, is almost completely absorbed in the counter. Also, the energy difference to the incident of 0.85 keV and the fluorescence yield are very small.



Saturday, 24 August 2013

X RAY FLUORESCENCE

INTRODUCTION

X-ray fluorescence (XRF) analysis is one of the most common non-destructive methods for  qualitative as well as quantitative determination of elemental composition of materials. It is suitable for solids, liquids as well as powders. There are two main methodological techniques that are wavelength dispersive analysis (WD-XRF) and energy dispersive analysis (ED-XRF) (In the next post we will briefly discuss about WDXRF & EDXRF ,this post will only explain the basics of x-ray fluorescence which is required to understand the upcoming posts about WDXRF & EDXRF ). The spectra are collected simultaneously in a wide energy range. The range of detectable materials covers all elements from Sodium (Na) to Uranium (U) and the concentration can range from 100% down to ppm. Detection limit depends upon the specific element and the sample matrix but in general heavier elements have higher detection limit.

X-ray Fluorescence (XRF) Spectroscopy involves measuring the intensity of X-rays emitted from a specimen as a function of energy or wavelength. The energies of large intensity lines are characteristic of atoms of the specimen. The intensities of observed lines for a given atom vary as the amount of that atom present in the specimen. Qualitative analysis involves identifying atoms present in a specimen by associating observed characteristic lines with their atoms. Quantitative analysis involves determining the amount of each atom present in the specimen from the intensity of measured characteristic X-ray lines. The emission of characteristic atomic X-ray photons occurs when a vacancy in an inner electron state is formed, and an outer orbit electron makes a transition to that vacant state. The  energy of the emitted photon is equal to the difference in electron energy levels of the transition. As the electron energy levels are characteristic of the atom, the energy of the emitted photon is characteristic of the atom. Molecular bonds generally occur between outer electrons of a molecule leaving inner electron states unperturbed. As X-ray fluorescence involves transitions to inner electron states, the energy of characteristic X-ray radiation is usually unaffected by molecular chemistry. This makes XRF a powerful tool of chemical analysis in all kinds of materials. In a liquid, fluoresced X-rays are usually little affected by other atoms in the liquid and line intensities are usually directly proportional to the amount of that atom present in the liquid. In a solid, atoms of the specimen both absorb and enhance characteristic X-ray radiation. These interactions are termed 'matrix effects' and much of quantitative analysis with XRF spectroscopy is concerned with correcting for these effects.

X rays are electromagnetic radiation. All X-rays represent a very energetic portion of

the electromagnetic spectrum (Table 1) and have short wavelengths of about 0.1 to 100 angstroms (Å). They are bounded by ultraviolet light at long wavelengths and gamma rays at short wavelengths X-rays in the range from 50 to 100 Å are termed soft X-rays because they have lower energies and are easily absorbed.The range of interest for X-ray is approximately from 0.1 to 100 Å. Although,angstroms are used throughout these notes, they are not accepted as SI unit. Wavelengths should be expressed in nanometers (nm), which are 10-9 meters (1 Å = 10-10 m), but most texts and articles on micro probe analysis retain the use of the angstroms. Another commonly used unit is the micron, which more correctly should be termed  micrometer  (μm), a micrometer is 104 Å. The relationship between the wavelength of electromagnetic radiation and its corpuscular energy (E) is derived as follows. 


For all electromagnetic radiation:

E = h ν ;

where:
h is the Planck constant (6.62 10-24 J.s);
ν is the frequency expressed in Hertz.

For all wavelengths,
ν = c / λ ;
where:
c = speed of light (2.99782 108 m/s);
λ= wavelength (Å).
Thus:
E = hc / λ = 1.9863610−24 /λ ;

where E is in Joule and λ in meters.
The conversion to angstroms and electron volts (1 eV = 1.6021 10-19 Joule) yields the
Duane-Hunt equation:
E(eV) 12.396/ (A)
= λ . 

Note the inversion relationship. Short wavelengths correspond to high energies and long wavelengths to low energies. Energies for the range of X-ray wavelengths are 124 keV (0.1 Å) to 124 eV (100 Å). The magnitudes of X-ray energies suggested to early workers that Xrays are produced from within an atom. Those produced from a material consist of two distinct superimposed components: continuum (or white) radiation, which has a continuous distribution of intensities over all wavelengths, and characteristic radiation, which occurs as a peak of variable intensity at discrete wavelengths.

PROPERTIES OF X-RAYS

A general summary of the properties of X-rays is presented below:
  • Invisible.
  • Propagate with velocity of light (3.10^8 m/s).
  • Unaffected by electrical and magnetic fields.
  • Differentially absorbed in passing through matter of varying composition, density and thickness.
  • Reflected, diffracted, refracted and polarized.
  • Capable of ionizing gases.
  • Capable of affecting electrical properties of solids and liquids.
  • Capable of blackening a photographic plate.
  • Able to liberate photo electron. And recoils electrons.
  • Emitted in a continuous spectrum.
  • Emitted also with a line spectrum characteristic of the chemical element.
  • Found to have absorption spectra characteristic of the chemical element.

THE ORIGIN OF X-RAYS

An electron can be ejected from its atomic orbital by the absorption of a light wave

(photon) of sufficient energy. The energy of the photon (hν) must be greater than the energy with which the electron is bound to the nucleus of the atom. When an inner orbital electron is ejected from an atom, an electron from a higher energy level orbital will transfer into the vacant lower energy orbital (Figure). During this transition a photon may be emitted from the atom. To understand the processes in the atomic shell, we must take a look at the Bohr’s atomic model. The energy of the emitted photon will be equal to the difference in energies between the two orbitals occupied by the electron making the transition. Due to the fact that the energy difference between two specific orbital shells, in a given element, is always the same (i.e., characteristic of a particular element), the photon emitted when an electron moves between these two levels will always have the same energy. Therefore, by determining the energy (wavelength) of the X-ray light (photons) emitted by a particular element, it is possible to determine the identity of that element.

PRINCIPLE OF THE X-RAY FLUORESCENCE PROCESS

If the primary energy of X-rays is equal to or is larger than the binding energy of an inner shell electron it is likely that electrons will be ejected and consequently vacancies are created. The hole state has certain life time and becomes refilled again. The transition of the excited atom into a state with lower energy occurs via two competitive processes, the above mentioned photoelectric and Auger effects. In the photoelectric effect, the recombination is accompanied by a transfer of electrons from the outer shells with energy Em into the inner shells with energy En filling the vacancies. This process induces the emission of a characteristic X-ray (fluorescence) photon with energy

                                                                   hV = Em - En

Therefore the energy of these secondary X-rays is the difference between the binding energies of the corresponding shells in the figure below. The excited atom can also recombine by emission of Auger electrons, instead of characteristic X-rays, via the Auger effect.
The probability that characteristic X-rays will be emitted - and not an Auger electron- varies from one element to another and is described as the fluorescence yield. For elements of low atomic numbers, the Auger effect dominates, whereas emission of characteristic X-rays is more likely for heavy elements.
Each element has its unique characteristic energy spectrum (Fluorescence spectrum) composed by the allowed transitions of the specific atom in the result of X-ray excitation. XRF technique consists on the study of the produced characteristic spectrum. The XRF emission induced by photoelectron effect is shown in figure below for an atom of titanium (Z=22), whose K-shell electron acquires sufficient energy to escape from the atom.




Photoelectric effect on the K-shell
An electron in the K-shell absorbs a
photon of the primary x-ray beam and
becomes free, while the atom gains a
vacancy in the K-shell.










The K lines production
An electron from the L or M shell “jumps in” to fill the vacancy and in  turn, produces a vacancy in the L or M shell. In the process, the atom emits a characteristic photon from the x-ray  range of electromagnetic spectrum, unique to this chemical element









The L lines production
After a vacancy is created in the L shell by either the primary beam photon or by the previous event, an electron from the M or N shell “jumps in” to occupy the vacancy. In this process, the atom emits a characteristic photon, unique to this chemical element, and a vacancy in the M or N shell is produced





  •  Ionization of the K-shell electron in the atom of Ti by photoelectric effect and emission of characteristic photons of different spectral series as a result of electron transitions in the atom.

              Electron transitions and emitted spectral lines in the atom after the K-shell ionization

X-ray fluorescence provides a rapid non-destructive means for both qualitative and quantitative analysis. A wide range of materials varying in size and shape can be studied with minimal requirements for sample preparation. Detection sensitivities as low as one part in a million can be obtained with this technique.The two types of X-ray fluorescence i.e EDXRF and WDXRF will be discussed and explained in the next post .

Some basic terms and definitions related to X- ray fluorescence, which can be useful in the upcoming post .

  • Attenuation coefficient – a natural logarithm of the ratio of the emergent and incident radiation intensities I / I0 divided by either the depth of the radiation penetration (linear attenuation coefficient) or the surface density (mass absorption coefficient).
  • Bremsstrahlung – a continuous spectrum produced by a charged particle moving with deceleration.
  • Continuous spectrum – a spectrum formed by photons with non-quantized energies in a wide range.
  • Detection limit – a lowest amount of chemical element that can be found with probability of 99%.
  • Detector resolution – possibility to distinguish two overlapping peaks in the spectrum; depends on the ratio of the distance between the two peaks and FWHM; usually accepted as a value of FWHM.
  • Efficiency of a detector – the ratio of the number of photons participated in creation of a useful signal in the detector to the total number of photons incident on the detector surface.
  • Energy-dispersive technique – the technique used to simultaneously detect the photons of the line spectrum in a wide range of energies.
  • Fluorescence – emission of photons by a substance that has absorbed photons with higher energy.
  • FWHM full width at half maximum of the peak usually measured in electronvolts.
  • Ionizing radiation – the particles or electromagnetic waves whose energy is sufficient to ionize a neutral atom or a molecule.
  • Line spectrum – a spectrum formed by photons with specific quantized energies only.
  • Matrix effects – The combined effect of all components of the sample other than the analyte on the measurement of the quantity of the analyte. The two main matrix effects are::

                           -(a) The attenuation of characteristic peak intensity due to inelastic scattering of photons, emitted by atoms of one chemical element, on atoms and electrons of other components
                            -(b)The enhancement of characteristic peak intensity due to additional excitation of atoms of one element by photons, emitted by other components.

  • Peak intensity – the value proportional to the total number of photons with same energy registered by a spectrometer and exposed as a bell-shaped curve called the peak.
  • Quantitative analysis – determination of amount of each component (chemical element) of a sample.
  • Spectral series – series of spectral peaks produced by electron transitions from different energy levels to one specific energy level; K-series corresponds to all transitions to the lowest possible energy level.
  • Spectrum – a function of a number of photons versus their energy, or versus their wavelength.
  • Spectrum background – A component of a spectrum which does not belong to the peak of interest, may be formed by bremsstrahlung radiation or by the tails of adjacent peaks.
  • X-ray tube – A kind of a vacuum tube with a filament as a cathode, emitting electrons, and a pure metal plate as an anode, producing radiation in the x-ray range of electromagnetic spectrum.