Atomic force microscopy (AFM) or scanning force microscopy (SFM) is a very high-resolution type of scanning probe
microscopy, with demonstrated resolution on the order
of fractions of a nanometer, AFM provides a 3D profile of the surface on a nanoscale, by
measuring forces between a sharp probe (<10 nm) and surface at very
short distance (0.2-10 nm probe-sample separation). The probe is supported on a
flexible cantilever. The AFM tip “gently” touches the surface and records the
small force between the probe and the surface.
The AFM
measures the forces acting between a fine tip and a sample. The tip is attached to the
free end of a cantilever, The cantilever is typically silicon or silicon
nitride with a tip radius
of curvature on the order of nanometres and it is brought very close to a surface. Attractive or repulsive
forces resulting from interactions between the tip and the surface will cause a
positive or negative bending of the cantilever. The bending is detected by
means of a laser beam, which is reflected from the back side of the cantilever.
The figure shows the basic concept of AFM :
What types of forces are measured?
Plot of force as a function of probe-sample separation |
The dominant interactions
at short probe-sample distances in the AFM are Van der Waals interactions.
However long-range interactions (i.e. capillary, electrostatic, magnetic) are
significant further away from the surface. These are important in other SPM
methods of analysis. During contact with the sample, the probe predominately
experiences repulsive Van der Waals forces (contact mode). This leads to
the tip deflection described previously. As the tip moves further away from the
surface attractive Van der Waals forces are dominant (non-contact mode).
Modes
of operation
CONTACT MODE
In the
So-called contact-AFM mode, the tip makes soft “physical contact” with the surface
of the sample. The deflection of the cantilever Dx is proportional to the force
acting on the tip, via Hook’s law, F=-k. x, where k is the spring constant of the
cantilever. In contact-mode the tip either scans at a constant small height above
the surface or under the conditions of a constant force. In the constant height
mode the height of the tip is fixed, whereas in the constant-force mode the deflection
of the cantilever is fixed and the motion of the scanner in z-direction is recorded.
By using contact-mode AFM, even “atomic resolution” images are obtained. For
contact mode AFM imaging, it is necessary to have a cantilever which is soft enough
to be deflected by very small forces and has a high enough resonant frequency to not be susceptible to vibration instabilities. Silicon Nitride tips are
used
for contact mode. In these tips, there are 4 cantilever with different geometries attached to each substrate, resulting in 4 different spring constants.
Probe with four different cantilevers with
different spring constants (N/m |
Advantages:
- High scan speeds
- Atomic resolution” is possible
- Easier scanning of rough samples with extreme changes in vertical topography
Disadvantages:
- Lateral forces can distort the image.
- Capillary forces from a fluid layer can cause large forces normal to the tip sample interaction
- Combination of these forces reduces spatial resolution and can cause damage to soft samples.
INTERMITTENT MODE (TAPPING)
The imaging is similar to contact.
However, in this mode the cantilever is oscillated at its resonant frequency,
Figure 4. The probe lightly “taps” on the sample surface during scanning,
contacting the surface at the bottom of its swing. By maintaining a constant
oscillation amplitude a constant tip-sample interaction is maintained and an image of the surface is
obtained.
Advantages:
- Higher lateral resolution (1 nm to 5 nm).
- Lower forces and less damage to soft samples in air.
- Almost no lateral forces.
Disadvantage:
- Slower scan speed than in contact mode.
NON CONTACT MODE
(Attractive VdW) The probe does not contact the sample
surface, but oscillates above the adsorbed fluid layer on the surface during
scanning. (Note: all samples unless in a controlled UHV or environmental
chamber have some liquid adsorbed on the surface). Using a feedback loop to
monitor changes in the amplitude due to attractive VdW forces the surface
topography can be measured.
Advantage:
- VERY low force exerted on the sample(10-12 N), extended probe lifetime.
Disadvantages:
- Lower lateral resolution, limited by tip-sample separation.
- Contaminant layer on surface can interfere with oscillation; usually need ultra-high vacuum (UHV) to have best imaging.
- Slower scan speed to avoid contact with fluid layer.
- Usually only applicable in extremely hydrophobic samples with a minimal fluid layer.
Applications :
The
number of applications for AFM has exploded since it was invented in 1986 and
Nowadays this technique is involved in many fields of Nanoscience and nanotechnology.
The remarkable feature of STM and AFM instruments is their ability
to examine samples not only in an ultrahigh vacuum but also on ambient conditions
or even in liquids. AFM can image the non-conducting surfaces, and
therefore it is very suitable for biological systems.
Possible applications of AFM are :
- Substrate roughness analysis.
- Step formation in thin film epitaxial deposition.
- Pinholes formation or other defects in oxides growth.
- Grain size analysis.
- Phase mode is very sensitive to variations in material properties, including surface stiffness, elasticity and adhesion.
- Comparing the tip-samples forces curves for materials to study the ratio of Young´s Modulus (graphite as a reference for measure of the indentation).
- Obtaining information of what is happening under indentation at very small loads .
- By In situ AFM analysis with changes in temperature we can study changes in the structure.
References
:
1. Basic Theory Atomic Force Microscopy (AFM) by Robert A. Wilson and Heather A. Bullen,*
Department of Chemistry, Northern Kentucky University, Highland Heights
2. Principles
of atomic force microscopy (AFM) written by Arantxa
Vilalta-Clemente , Aristotle University, Thessaloniki, Greece and Kathrin Gloystein, Aristotle University,
Thessaloniki, Greece
Amazing blog. Thanks for sharing this wonderful information with us
ReplyDeleteAtomic Force Microscope
Thanks so much for a great post. I'd like to know more about these topics and hope that I can receive more insight into this topic.
ReplyDeleteFor more information visit here: Atomic Force Microscope
Hey, Nice blog and knowledgeable content.
ReplyDeleteAtomic Force Microscope
Informative blog thank for sharing with us
ReplyDeleteSpring contact probe